28D ACS® BIO Lens VOLK
Primary Application - Industry Standard Diagnostic Lens in an Autoclavable Format
The quantity in the package | 1 couples |
Shipping price | The Lack Of |
The Availability Of | na zamówienie 1 couples |
Pełna oferta Volk Optical USA | |
The bar code | |
EAN |
PART #V28LCACSPV
Primary Application - Fundus Scanning Lens in an Autoclavable Format
- Steam sterilizable for use in a surgical environment
- High quality Permaview® glass withstands the rigors of repeated sterilization
- Excellent for small pupil diagnosis and treatment
Field of View | Image Magnification | Laser Spot Magnification | Working Distance |
53° / 69° | 2.27x | 0.44x | 33 mm |
Safety Information
BIO ACS LENSES:
Do not use a microfiber cloth, as over time these tend to collect dirt and dust which can damage the anti reflective coating on the lens.
Never use a device that shows any sign(s) of damage.
Failure to follow the point-of-use processing steps could adversely affect further decontamination steps.
Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the member state in which the user and/or patient is established.
A thorough, manual cleaning process is recommended. Corrosive cleaning agents (i.e. acids, alkalines, etc) are not recommended. Detergent cleaning agents with neutral ph are recommended.
Do not use detergents that contain any type of emollients.
Wiping the lens with a microfiber cloth will cause lens damage. Be sure to use only a soft, lint-free
cotton cloth.
Ensure the device is completely submerged in the disinfectant solution for the entirety of the recommended or desired soak time. Do not allow the device to become unsubmerged from the disinfectant solution. Exposure to disinfectant solutions beyond the recommended soak time, and/or exposure to higher concentrations of disinfectant solution, will result in accelerated degradation of most Volk product.
Colored rings may discolor when exposed to Sodium Hypochlorite or Glutaraldehyde. To avoid further degradation please follow only the disinfection procedures indicated for these products in instruction. This color change is purely cosmetic and will not affect the function of the lens.
Do not brush lens portion to avoid scratching. Use a soft, cotton cloth.
Do not sterilize lenses within standard (black leatherette) lens cases as they are not meant for use in sterilization systems.
For use by trained personnel only.
Manufacturer details
Volk Optical INC.
7893 Enterprise Dr, Mentor
OH 44060 Mentor,
United States
+1 440-942-6161
volk@volk.com
Responsible person
Rudolf Riester GmbH
Bruckstrasse, 31
72417 Jungingen
Germany
+49747792700
pfister@riester.de